How much can radiation dose vary between hospitals performing similar computed tomography examinations, and what does this mean for patient safety?
This study evaluated computed tomography dose-length product, DLP, across ten hospitals in Abuja, Nigeria, to assess variations in radiation exposure and identify opportunities for improving dose optimization.
The researchers used polymethyl methacrylate phantoms, a 100 mm pencil ionization chamber, an electrometer, and a CT probe to measure radiation doses for chest and head examinations.
DLP was selected because it combines the volume CT dose index with scan length and therefore provides a practical indicator of the total radiation output associated with a CT examination.
The results showed substantial differences in radiation dose across the ten facilities.
For chest examinations, measured DLP values ranged from 79.36 卤 0.001 mGy路cm at Hospital H to 677.40 卤 0.01 mGy路cm at Hospital D.
For head examinations, values ranged from 230.14 卤 0.004 mGy路cm at Hospital A to 1165.73 卤 0.3 mGy路cm at Hospital I.
Most hospitals recorded DLP values below the European diagnostic reference levels used in the study.
However, Hospital D exceeded the chest reference level of 650 mGy路cm, while Hospitals E and I exceeded the head reference level of 1050 mGy路cm.
These differences point to considerable variation in scanning practices and equipment performance between facilities.
The study suggests that factors such as scanner generation, acquisition protocols, slice thickness, number of slices, equipment calibration, and operator-selected settings may contribute to differences in dose.
Older CT systems may also produce higher dose-length products because of limitations in beam geometry and dose-reduction technology, while newer multidetector scanners may offer more efficient exposure control.
The findings further indicate that non-standardized scanning protocols can result in patients receiving substantially different radiation doses for comparable examinations at different hospitals.
This is particularly important because CT examinations generally involve higher radiation doses than conventional radiography, making dose optimization essential for reducing unnecessary exposure.
Overall, the study highlights the need to strengthen radiation-safety practices across CT facilities in Abuja.
The authors recommend routine quality-control programmes, regular equipment calibration, standardized scanning protocols, and refresher training for CT operators.
They also emphasize the importance of applying the As Low as Reasonably Achievable, ALARA, principle so that diagnostic image quality is maintained while unnecessary radiation exposure is minimized.
The findings could also contribute to the development of more consistent local diagnostic reference levels and improved monitoring of CT radiation doses in Nigeria.